Calibration slide for stereomicroscopes.
The slide has a 10x10mm reticle etched on it. In the center of the slide there are crosshairs – two micrometer scales of 200 divisions. The scale division is 0.05mm and the length of each scale is 10mm.
The calibration slide (stage micrometer) is used to determine the magnification linear field of view of the microscope and the division value of the eyepiece scales and reticles.
With a digital camera the slide is used to calibrate the image analysis software for each objective.

Calibrating the image analysis software

The calibration is performed for each objective that will be used in observations. The calibration slide is used to determine the calibration factor for the image analysis software.

Calibration procedure:

  1. Fix the calibration slide on the microscope stage.
  2. Select the desired objective and set the maximum camera resolution.
  3. Get a contrast image of the scale on the monitor screen and capture the image.
  4. Select “Calibration” mode in the image analysis software.
  5. Click with the mouse to define the maximum visible distance between the first and last divisions of the scale.
  6. Enter the resulting scale size in real units into the image analysis software.
  7. Enter the calibration name by specifying the current objective magnification.
  8. Check the result: In “Measurement” mode measure a few divisions of the calibration slide scales in real units.
  9. The image analysis software will save the calibration factor under the selected name. After that it will automatically convert the results to the selected units of measurement.

Calibration of the eyepiece scale

The calibration of the eyepiece scale is performed for each objective that will be used in observations. The slide should be placed on the stage and the microscope should be focused on the scale. The eyepiece should be rotated in the eyepiece tube until the scales of a slide and an eyepiece are aligned parallel to each other.

Calculation for each objective is carried out according to the formula:
Е=ТL/A where
E – eyepiece division value
Т – stage division value specified on the stage micrometer (0.05mm)
L – number of stage micrometer divisions
А – number of eyepiece scale divisions.

The resulting values of the eyepiece scale division for each objective shall be recorded.

When measuring an object first you should count the number of eyepiece scale divisions that the object covers. The result is multiplied by the scale value previously calculated for the selected objective. The resulting value will be the true linear value of the object.

Brand MAGUS
Warranty 2 years
EAN 5905555017211
Package size (LxWxH) 4x3x1 cm
Shipping Weight 0.01 kg
Material optical glass
Scale length mm 10
Number of scale divisions 200
Scale value mm 0.05
Reticle size mm 10х10
Number of reticle cells in a row 20
Reticle cell size mm 0.5х0.5
Calibration slide size mm 24

Returns Policy

We have a 14-day return policy. If you are not satisfied with your purchase, please return it in its original packaging and be sure to put the original product packaging inside the shipping box to avoid any damage..

All items must be in their original condition and unused. The value of your purchase will be refunded. .

Shipping charges will not be refunded.

If the return is due to a defective product, you may be compensated for the return shipping.

Please contact us for more details if you need to make a return..

SHIPPING POLICY

- From 0 to 2 kg - shipping cost: 12,50 euros.

- From 2.001 kg to 30 kg - shipping cost: 17,00 euros.

- From 30,001 kg - shipping cost: 30,00 euros.

In case of collection at our warehouse or resale points - the shipping is free.

Related Products