The Ermenrich Zing DZ40 is a versatile measuring tool for power engineers, instrumentation and control (I&C) technicians, and industrial maintenance professionals. The heavy-duty magnetic core with a 30mm jaw opening safely accommodates thick power cables, allowing for the testing of high-load mains without interrupting production processes or breaking the circuit.

High-current measurement

The key feature of the Ermenrich Zing DZ40 is its ability to measure AC and DC currents up to 1,000A. The device is ideal for inspecting electrical panels, evaluating the power consumption of industrial machinery, welding stations, and heavy-duty electric motors. DC current measurement is indispensable for monitoring industrial UPS systems and large battery banks. The 600V voltage limit fully covers the requirements for monitoring 230/400V three-phase networks.

Temperature measurement and operating conditions

The device measures temperatures ranging from –20°C to +1,000°C using a thermocouple. This is essential for auditing drying chambers, boiler plants, industrial furnaces, and internal combustion engine (ICE) exhaust systems. The clamp meter is built for harsh environments, operating reliably at ambient temperatures from –10°C to +50°C. This allows for measurements in unheated workshops, hangars, or outdoors during transitional seasons.

Frequency analysis and component testing

Despite its heavy-duty focus, the device handles electronic component diagnostics with ease. It can measure frequencies up to 10MHz and pulse duty cycles, which is useful for verifying control and PWM signals. The capacitance measurement mode (up to 100µF) helps test capacitors and identify dried-out components. An audible buzzer, which triggers at resistances below 100Ω, is provided for locating open circuits. Moreover, the semiconductor test function, featuring an open-circuit voltage of 1.5V, is ideal for testing rectifier bridges and power diodes.

Reliability and safety standards

The tool complies with the CAT III 600V overvoltage category, making it suitable for working in electrical panels and with fixed building wiring. Readings are displayed on a 4,000-count screen with a bright backlight that is invaluable when working in poorly lit electrical rooms and workshops. The device is powered by a single 9V battery (not included), and an auto-off function shuts down the meter after prolonged inactivity.

Key features:

  • AC and DC current measurement up to 1,000A is suitable for diagnosing energy-intensive installations
  • 30mm jaw opening accommodates thick cables and busbars
  • Temperature monitoring up to +1,000°C; use the thermocouple to audit furnaces, boilers, and engines
  • Frequency measurement up to 10MHz is useful for configuring control signals
  • CAT III 600V overvoltage category ensures safe operation in electrical panels
  • Wide operating temperature range (–10°C to +50°C) provides accurate measurements outdoors and in harsh environments

The kit includes:

  • Digital clamp meter
  • Probe – 2 pcs.
  • Thermocouple
  • Pouch
  • User manual and warranty card
Brand Ermenrich
Warranty 5 years
EAN 5905555028705
Package size (LxWxH) 24x14x6 cm
Shipping Weight 0.71 kg
Selection of the measuring limit automatic, manual
Basic functions ammeter, capacitance meter, duty cycle meter, frequency meter, ohmmeter, thermocouple, voltmeter
DC voltage measuring limit 400mV: accuracy ±(0.8%+3); 4V/40V/400V: accuracy ±(1.5%+3); 600V: accuracy ±(2.0%+3)
AC voltage measuring limit 400mV: accuracy ±(0.8%+20); 4V/40V/400V: accuracy ±(1.8%+5); 600V: accuracy ±(1.0%+4)
DC current measuring limit 40A: accuracy ±(2.8%+10); 400A: accuracy ±(2.8%+5); 1,000A: accuracy ±(3.0%+5)
AC current measuring limit 40A: accuracy ±(3.0%+10); 400A: accuracy ±(3.0%+5); 1,000A: accuracy ±(3.0%+5)
Resistance measuring limit 400Ω: accuracy ±(1.0%+4); 4kΩ/40kΩ/400kΩ: accuracy ±(1.5%+2); 4MΩ: accuracy ±(2.5%+3); 40MΩ: accuracy ±(3.5%+5)
Capacity measuring limit 40nF: accuracy ±(5.0%+100); 400nF: accuracy ±(3.0%+5); 4µF/40µF: accuracy ±(3.5%+5); 100µF: accuracy ±(5.0%+5)
Frequency measurement limits 5Hz/50Hz/500Hz/5kHz/50kHz/500kHz/10MHz: accuracy ±(1.2%+2)
Duty cycle measurement range 0.5–99.0%: accuracy ±(1.2%+2)
Temperature measurement range –20...+1,000°C: accuracy ±(3%+5°C)
Wire mapping yes
Continuity test +, threshold <100Ω, test current <1mA
Diodes test yes
Diode test +, test current 0.3mA, open-circuit voltage 1.5V DC
Hold function yes
DCA Zero yes
Input impedance 7.8MΩ (DC and AC)
Throughput capacitance 50/60Hz (AAC/VAC)
Display yes
Display counts 4000
Display backlight yes
Safety standards EN 61010-1:2010+A1:2019; EN IEC 61010-2-032:2021+A11:2021
Overvoltage category CAT III 600V
Operating temperature range, °C -10 — 50
Power supply 9V battery (not included)
Auto power off yes

Returns Policy

We have a 14-day return policy. If you are not satisfied with your purchase, please return it in its original packaging and be sure to put the original product packaging inside the shipping box to avoid any damage..

All items must be in their original condition and unused. The value of your purchase will be refunded. .

Shipping charges will not be refunded.

If the return is due to a defective product, you may be compensated for the return shipping.

Please contact us for more details if you need to make a return..

SHIPPING POLICY

- From 0 to 2 kg - shipping cost: 12,50 euros.

- From 2.001 kg to 30 kg - shipping cost: 17,00 euros.

- From 30,001 kg - shipping cost: 30,00 euros.

In case of collection at our warehouse or resale points - the shipping is free.

Related Products